Proline Promag 50W, 53W
14 Endress+Hauser
Operating conditions
Installation conditions
Installation instructions Mounting location
Correct measuring is possible only if the pipe is full. Avoid the following locations:
Highest point of a pipeline. Risk of air accumulating
Directly upstream of a free pipe outlet in a vertical pipe.
F06-5xxxxxxx-11-00-00-xx-000
Installation of pumps
Do not install the sensor on the intake side of a pump. This precaution is to avoid low pressure and the conse-
quent risk of damage to the lining of the measuring tube. Information on the lining's resistance to partial vac-
uum can be found on Page 20.
It might be necessary to install pulse dampers in systems incorporating reciprocating, diaphragm or peristaltic
pumps. Information on the measuring system's resistance to vibration and shock can be found on Page 20.