Promag CON-485 Series Informações Técnicas Página 33

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 40
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 32
Proline Promag 50W, 53W
Endress+Hauser 33
Material load diagrams Caution!
The following diagrams contain material load curves (reference curves) for various process connections relating
to the fluid temperature. But the maximal permissible fluid temperature always depends on the lining material
of the sensor and/or of the sealing material (Page 20).
Flange connection to EN 1092-1 (DIN 2501)
Material: RSt37-2 (S235JRG2) / C22 / FE 410W B
F06-xxFxxxxx-05-xx-xx-xx-000
Flange connection to EN 1092-1 (DIN 2501)
Material: 316L / 1.4571
F06-xxFxxxxx-05-xx-xx-xx-001
0
5
10
15
20
25
PN 25
PN 16
PN 10
PN 6
35
30
40
PN 40
[bar]
-60
-40
-20
020
40 60 80 100 120 140 160 180
[°C]
0
5
10
15
20
25
PN 25
PN 16
PN 10
35
30
40
PN 40
[bar]
-60
-40
-20
020
40 60 80 100 120 140 160 180
[°C]