Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning
Endress+Hauser 23
5 Operation / adjusting device settings -
Set-up to commissioning
5.1 Basics on set-up directly at the device
• The function of the operating keys is described in the fields directly above the corresponding keys
on the screen. Empty fields mean that the corresponding keys currently have no function.
• Access to the set-up is released ex works and can be locked in various ways:
– Entering a 4-digit release code (see Chapter 5.7.1 “Basic settings”)
– Clear password / ID combinations for administrator and user (see Chapter 5.7.6 “Administra-
tor”)
– Activating a digital input (see Chapter “Signal settings - Digital inputs”)
– Attaching the rear panel cover with leaded sealing.
• When locked, basic settings can be checked. Users can not make any changes. An administrator
however, who is defined during commissioning, can make the following changes:
– Adding or deleting new or already registered users (e.g. new employees or employees who have
left the company)
– Adding, changing or deleting texts / comments. Comments that have already been saved are
not affected by this
"
Caution!
After changing settings that affect the measured value memory (recognisable by the
“ * ” symbol in the header), the measured value memory is deleted. If the previous measured values
are still required, please update the ATA flash card and take these values, or read out the device
using an interface. Afterwards, end the set-up. When the new set-up data is adopted, the old measu-
rement data in the memory and on the ATA flash card is deleted, the device is reset and restarted.
5.2 Recommendation for commissioning and safe access
protection:
1. Set the corresponding parameters or load a parameter set onto the device.
2. Enter a release code (or activate a control input in order to lock access to the set-up), start and
check that the device is functioning properly.
3. After the operational test has been successfully completed, register an administrator and a user
if required (see “Administration”).
!
Note!
There always has to be an administrator registered before he himself can define users.
4. Attach the rear panel cover and seal it. This prevents any further change to the device settings.
Use of the interface(s) is thereby only possible in read-only mode.
5.3 Access rights:
!
Note!
The availability of the rear panel cover is monitored with the help of a light barrier. This monitoring
can be activated/deactivated using the device software. Depending on the setting, the administrator
or the user has access rights to the set-up, in order to change the settings. When locked, the settings
in the set-up can only be read:
In principle, the device settings can only be read. The set-up can only be changed under
certain conditions (see table).